Performance:
- Intuitive, easy use with direct
function keys
- 12-channel Resting ECG
- Measurements
- Interpretation
- Serial data transfer to PC for storage of
both Resting ECG and Spirometry
recordings
- XML data transfer for Resting ECG to
PC or Information Systems via Ethernet
or modem interfaces (optional)
- A4-formated (8.5 x11 inches) printouts
on internal or external printer
- Basic Exercise test capability (optional)
- Possibility for external monitor
connection (optional)
Resting ECG
Operating elements:
- Integrated 3-channel monitor
- Alphanumeric keyboard
- Direct keys for the most important functions
- Easy Print: Integrated high-resolution
A4 (8.5 x11 inches) thermal printer,
subsequent printout on thermoreactive
paper possible
- Interface for external printer for A4
(8.5 x11 inches) printouts on regular pape
- • Internal memory for up to
40 records
- Interface for data-transmission
Exercise ECG
Interface:
- RS-232 port for peripheral devices such
as treadmills,bicycles etc.
Protocols:
- For exercise ECG, several stand-ard
protocols are available for bicycle
and treadmill ergometers; the user
can either select a standard protocol
from a table or define a new protocol
- J-point settings: J+20,+40,+60,+80 ms;
default: +60 ms
Display on the LCD:
- 3 leads can be selected via direct keys
- Heart rate
- ST slope and amplitude (in mV/mV/s)
of a lead (the lead can be selected in
the set-up; default is V5)
- Stage,time,and current load
- Blood pressure (Blood pressure
measurement: Automatic – if a bicycle
contains an interface for the
blood pressure transfer – or manual)
Printouts:
- Online printout: 6 leads on the internal
thermal printer
- 10-sec.printout: A 10-sec.printout can
be initiated at any time on the internal
thermal printer
- Two print formats are available for
stage printouts on the internal thermal
printer:
– 2*6 25 mm/s and 50 mm/s,1 page
– 4*3 + 1 rhythm,25 mm/s (50 mm/s),
1 page
- Final report: All test data is listed in a tabular form
XML Data Transmission (optional)
- The AT-102 is available with the internal
SCHILLER Communication Module
(SCM),which transmits Resting ECG
data on XML format to a PC or
Information Systems via:
– Ethernet interface; or
– Built-in analogue modem (optional)
- This enables the AT-102 to be an
integrated part of a larger systems
solution
- Secure transmission of patient data
via the Internet is possible using
1024-bit strong encryption
- Equipped with the add-on SCM Server,
the SCHILLER Data
Management System SEMA-200
can receive the XML output from the
SCHILLER Communication Module
Spirometry:
- Unique SP-250 with disposable sensor
to reduce the risk of contamination
- SP-260 with reusable sensor
- Simple and fast calibration
- Memory for up to 40 records
- Inspiratory and exspiratory test; pre
and post medication tests
– Selectable predicted value tables
– Real-time curves and measurement
information on the large screen
Specifications
System:
Dimensions: 400 x 330 x 101 mm,approx. 5 kg
Built-in monitor: 120 x 90 mm effective display area,
320 x 240 dots resolution
On-screen status indicators:
Battery status, date, time,power source
Control panel and keyboard: User-friendly, alphanumeric
keyboard, LED indicator and LC display
Power supply requirements: 220–240 V (nominal), 50/60 Hz;
110–115 V (nominal), 50/60 Hz; stand-alone operation with
built-in rechargeable battery; LED indicator for mains
operation, integrated power supply unit
Battery capacity: 4 hours of normal use (approx. 300
automatic ECG printouts or 100 spirometry records)
Power consumption: Max. 30 VA
Line frequency filter: Distortion-free suppression of superimposed
50 or 60 Hz sinusoidal interferences by means of
adaptive digital filtering (SCHILLER Powerline Filter SPF)
Frequency range of digital recording system: 0 Hz – 150 Hz
(IEC/AHA)
Chart paper: Thermo-reactive, Z-folded, 210 mm wide
(A4, 8.5 x 11 inches)
Printing process: High-resolution thermal head printer,8 dots/mm
(amplitude axis), 40 dots/mm (time axis) @ 25 mm/s
Interfaces: RS-232 interface to connect spirometry sensor,
data transmission to PC (SEMA-200) and external modem
connection; parallel port for external printer; optional
Ethernet interface and optional internal modem for XML data
transmission
Memory: Possibility to save up to 40 Resting ECG or Spirometry
measurements
Environmental conditions:
– Temperature, operating: 10° to 40°C
– Temperature, storage: -10° to 50°C
– Relative humidity: 25 to 95% (no condensation)
– Pressure during operation: 700 to 1060 hPa
Safety Standards:
Safety Standards: IEC/EN 60601-1; UL 60601-1;
C22.2 No. 601.1-M90; IEC/EN 60601-2-25;
IEC/EN 60601-1-2 (EMC)
Protection Class: I according to IEC/EN 60601-1 (with internal
power supply)
Applied Part: CF according IEC/EN 60601-1
Conformity: according Directive 93/42/EEC (Medical
Devices)
Classification: IIa according Directive 93/92/EEC
Technical Data for ECG:
Patient input circuit: Fully floating and isolated, defibrillationprotected
(only with original SCHILLER patient cable)
Monitor display:
– 3-channel display of the selected leads
– 25, 50 mm/s
– 5, 10,20 mm/mV
– Filter status (on/off)
– Insufficient electrode contact
– Heart Frequency, HF
– mm/mv,mm/s
Leads: 12 simultaneous leads: Standard / Cabrera
Chart printout speed: 5/10/25/50 mm/s (manual print)
Sensitivities: 5/10/20 mm/mV, either automatically adjusted or
manually selected
Automatic lead programs:
– 6/12-channel presentations of 12 simultaneously recorded
standard leads on one or more A4 (8.5 x 11 inches) pages
– Numerous printout formats can be selected
Data record:
– Patient data (name, age, height,weight, BP), user ID
– Listing of all ECG recording conditions (date, time, filter)
– Measurement program (M): ECG measurements results
(intervals, amplitudes, electrical axes), average complexes
with measurement reference markings
– Interpretation (C) with adult and pediatric guidance
Recording track: 6/12-channel presentation, optimal positioning
on a width of 200 mm,automatic baseline adjustment
Filter: Myogram filter (muscle tremor filter): 25 Hz or 35 Hz,
can be switched on/off (SCHILLER Baseline Stabilizer SBS ,
SCHILLER Smoothing Filter SSF)
ECG amplifier:
– Simultaneous recording of all 9 active electrode signals
(= 12 leads)
– Sampling frequency: 1000 Hz
– Pacemaker detection: 2 mV/ 0,1 ms
Technical Data for Spirometry (Option):
Measured values:
FVC: FVC, FEV0.5,FEV1.0, FEV3.0, FEV0.5/FVC, FEV1,0/FVC,
FEV3.0/FVC, FEF0.2-1.2, FEF25-75%, FEF75-85%, PEF, FEF25%, FEF50%, FEF75%,
FIVC, FIV1.0 , FIV1.0/FIVC, FIV1.0/FVC,PIF, FIF50%, FMFT
SVC: SVC, ERV, IRV,TV
MVV: MVV, RR,TV
Presentation possibilities (printout and screen):
– Flow/volume graph
– Volume/time graph
– Table of measured values
– Realtime flow curve
Data record:
– Patient data (name, age, height,weight), user ID
– Registration conditions (date, time, date of last calibration)
– Flow/volume graph and/or volume/time graph
– Table of measured values with PREDICTED/ACTUAL/
DIFFERENTIAL values
– Diagnosis guidance
– Memory for more than 60 ECGs or spirometry records
Prediction equation:
Adults: ECCS,Austria, Crapo, Morris, Knudson, Knudson76,
Polgar, Berglund, Finland, India, Composite
Children: Quanjer & Tammeling,Austria, India, Knudson,
Knudson76, Polgar
Extrapolated predicted values
Comparison of PRE/POST medication is possible
Standards Compliance: ATS, OSHA, NIOSH
SPIROVIT SP-250 Pneumotach Flowsensor for pulmonary
function testing with disposable mouthpiece:
Dimensions of SP-250: 118 x 36 x 28 mm,approx. 120 g;
4.6 x 1.4 x 1.1 in.,approx. 0.26 Ibs
Measuring method: Pneumotachometer
Measuring accuracy: According ATS < 3%
Flow impedance: < 0.2 mbar*s/l at 12 l/s
SPIROVIT SP-260 Pneumotach Flowsensor for pulmonary
function testing with reusable mouthpiece:
Dimensions of SP-260: 125 x 36 x 28 mm,approx. 160 g;
4.9 x 1.4 x 1.1 in.,approx. 0.34 Ibs
Measuring method, Measuring accuracy, Flow impedance
same as SP-250
Scope of delivery:
– 1 pneumotach sensor; choice of SP-250 disposable
(P/N 2.100022) or SP-260 reusable sensor (P/N 2.100551)
– Accessories: 2 noseclips and either 1 pack of disposable
plastic mouthpieces for SP-250 (P/N 2.100077) or 1 pack
of disposable filters for SP-260 (P/N 2.100123)
– 1 operating manual
Scope of delivery for CARDIOVIT AT-102 Standard:
CARDIOVIT AT-102 Resting ECG with 12 simultaneous leads,
pacemaker detection,automatic ECG measurement
Accessories:
– 1 10-lead patient cable
– 1 set of electrodes or disposable electrodes
– 1 power cable
– 1 pack chart paper
– Operating manual
Software options:
– Computer-aided ECG interpretation for pediatric and adult
ECGs (C)
– Basic exercise test
– SEMA-200 PC software to save, validate and archieve ECG
and spirometry data on PC
Hardware options:
– Equipment trolley
– Spirometry sensor SP-250 or SP-260
– Calibration syringe
– Vacuum electrode system VAC-100
– External printer
– External monitor connection
– Internal SCHILLER Communication Module with optional
built-in analogue modem
*Technical Data are subject to change without notice.