Please Wait...
Includes: Patient Lead Cable, Pack of Electrodes, Pack of Paper, Pack of Clips, Power Cable, Manual.
Because we are an authorized dealer, the manufacturer has asked us to not show our prices. To contact us regarding your guaranteed lowest price, please click the price quote button.
System: Dimensions: 400 x 330 x 101 mm,approx. 5 kg Built-in monitor: 120 x 90 mm effective display area, 320 x 240 dots resolution On-screen status indicators: Battery status, date, time,power source Control panel and keyboard: User-friendly, alphanumeric keyboard, LED indicator and LC display Power supply requirements: 220–240 V (nominal), 50/60 Hz; 110–115 V (nominal), 50/60 Hz; stand-alone operation with built-in rechargeable battery; LED indicator for mains operation, integrated power supply unit Battery capacity: 4 hours of normal use (approx. 300 automatic ECG printouts or 100 spirometry records) Power consumption: Max. 30 VA Line frequency filter: Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal interferences by means of adaptive digital filtering (SCHILLER Powerline Filter SPF) Frequency range of digital recording system: 0 Hz – 150 Hz (IEC/AHA) Chart paper: Thermo-reactive, Z-folded, 210 mm wide (A4, 8.5 x 11 inches) Printing process: High-resolution thermal head printer,8 dots/mm (amplitude axis), 40 dots/mm (time axis) @ 25 mm/s Interfaces: RS-232 interface to connect spirometry sensor, data transmission to PC (SEMA-200) and external modem connection; parallel port for external printer; optional Ethernet interface and optional internal modem for XML data transmission Memory: Possibility to save up to 40 Resting ECG or Spirometry measurements Environmental conditions: – Temperature, operating: 10° to 40°C – Temperature, storage: -10° to 50°C – Relative humidity: 25 to 95% (no condensation) – Pressure during operation: 700 to 1060 hPa
Safety Standards: Safety Standards: IEC/EN 60601-1; UL 60601-1; C22.2 No. 601.1-M90; IEC/EN 60601-2-25; IEC/EN 60601-1-2 (EMC) Protection Class: I according to IEC/EN 60601-1 (with internal power supply) Applied Part: CF according IEC/EN 60601-1 Conformity: according Directive 93/42/EEC (Medical Devices) Classification: IIa according Directive 93/92/EEC
Technical Data for ECG: Patient input circuit: Fully floating and isolated, defibrillationprotected (only with original SCHILLER patient cable) Monitor display: – 3-channel display of the selected leads – 25, 50 mm/s – 5, 10,20 mm/mV – Filter status (on/off) – Insufficient electrode contact – Heart Frequency, HF – mm/mv,mm/s Leads: 12 simultaneous leads: Standard / Cabrera Chart printout speed: 5/10/25/50 mm/s (manual print) Sensitivities: 5/10/20 mm/mV, either automatically adjusted or manually selected Automatic lead programs: – 6/12-channel presentations of 12 simultaneously recorded standard leads on one or more A4 (8.5 x 11 inches) pages – Numerous printout formats can be selected Data record: – Patient data (name, age, height,weight, BP), user ID – Listing of all ECG recording conditions (date, time, filter) – Measurement program (M): ECG measurements results (intervals, amplitudes, electrical axes), average complexes with measurement reference markings – Interpretation (C) with adult and pediatric guidance Recording track: 6/12-channel presentation, optimal positioning on a width of 200 mm,automatic baseline adjustment Filter: Myogram filter (muscle tremor filter): 25 Hz or 35 Hz, can be switched on/off (SCHILLER Baseline Stabilizer SBS , SCHILLER Smoothing Filter SSF) ECG amplifier: – Simultaneous recording of all 9 active electrode signals (= 12 leads) – Sampling frequency: 1000 Hz – Pacemaker detection: 2 mV/ 0,1 ms
Technical Data for Spirometry (Option): Measured values: FVC: FVC, FEV0.5,FEV1.0, FEV3.0, FEV0.5/FVC, FEV1,0/FVC, FEV3.0/FVC, FEF0.2-1.2, FEF25-75%, FEF75-85%, PEF, FEF25%, FEF50%, FEF75%, FIVC, FIV1.0 , FIV1.0/FIVC, FIV1.0/FVC,PIF, FIF50%, FMFT SVC: SVC, ERV, IRV,TV MVV: MVV, RR,TV Presentation possibilities (printout and screen): – Flow/volume graph – Volume/time graph – Table of measured values – Realtime flow curve Data record: – Patient data (name, age, height,weight), user ID – Registration conditions (date, time, date of last calibration) – Flow/volume graph and/or volume/time graph – Table of measured values with PREDICTED/ACTUAL/ DIFFERENTIAL values – Diagnosis guidance – Memory for more than 60 ECGs or spirometry records Prediction equation: Adults: ECCS,Austria, Crapo, Morris, Knudson, Knudson76, Polgar, Berglund, Finland, India, Composite Children: Quanjer & Tammeling,Austria, India, Knudson, Knudson76, Polgar Extrapolated predicted values Comparison of PRE/POST medication is possible Standards Compliance: ATS, OSHA, NIOSH
SPIROVIT SP-250 Pneumotach Flowsensor for pulmonary function testing with disposable mouthpiece: Dimensions of SP-250: 118 x 36 x 28 mm,approx. 120 g; 4.6 x 1.4 x 1.1 in.,approx. 0.26 Ibs Measuring method: Pneumotachometer Measuring accuracy: According ATS < 3% Flow impedance: < 0.2 mbar*s/l at 12 l/s SPIROVIT SP-260 Pneumotach Flowsensor for pulmonary function testing with reusable mouthpiece: Dimensions of SP-260: 125 x 36 x 28 mm,approx. 160 g; 4.9 x 1.4 x 1.1 in.,approx. 0.34 Ibs Measuring method, Measuring accuracy, Flow impedance same as SP-250 Scope of delivery: – 1 pneumotach sensor; choice of SP-250 disposable (P/N 2.100022) or SP-260 reusable sensor (P/N 2.100551) – Accessories: 2 noseclips and either 1 pack of disposable plastic mouthpieces for SP-250 (P/N 2.100077) or 1 pack of disposable filters for SP-260 (P/N 2.100123) – 1 operating manual
Scope of delivery for CARDIOVIT AT-102 Standard: CARDIOVIT AT-102 Resting ECG with 12 simultaneous leads, pacemaker detection,automatic ECG measurement Accessories: – 1 10-lead patient cable – 1 set of electrodes or disposable electrodes – 1 power cable – 1 pack chart paper – Operating manual Software options: – Computer-aided ECG interpretation for pediatric and adult ECGs (C) – Basic exercise test – SEMA-200 PC software to save, validate and archieve ECG and spirometry data on PC Hardware options: – Equipment trolley – Spirometry sensor SP-250 or SP-260 – Calibration syringe – Vacuum electrode system VAC-100 – External printer – External monitor connection – Internal SCHILLER Communication Module with optional built-in analogue modem
*Technical Data are subject to change without notice.