General
Dimensions: 9 3/8" x 9" x 3" (238 x 229 x 76 mm)
Weight: Less than 5 pounds (2.3 kg)
Operating Temperature: 32° to 122°F (0° to 50°C)
Storage Temperature (without battery): -22° to 150°F (-30° to 65°C)
Humidity: Up to 95% or 10° to 95°C (non-condensing)
Altitude: -500 to 15,000 feet (-150 to 4570 m)
Shock: Mil Std 810E method 516.4, procedure 1 (40G, 6-9ms pulse, 1/2 sine each axis)
Vibration: Mil Std 810E method 514.4, category 10
Water resistance: IEC 529 IPX4
Data Managment
Event Documentation: Internal and via MRL Datacard
Internal Memory Capacity: 1MB: 100 4-sec ECG samples or 300 time stamped events
Datacard Capacity:
> 4MB - 90 min of Continuous ECG - 40 min of ECG and audio
> 8MB - 3 hrs of Continuous ECG - 80 min of ECG and audio
> 16MB - 6 hrs of Continuous ECG - 2.5 hrs of ECG and audio
Playback: MRL PIC, PIC 2 or Smartview
Quick Report: Treatment Summary, Event Log, Test Log
Communication: Serial port via RS-232 to PC and Printer
Defibrillator
Output: Welch Allyn Biphasic technology
Auto/Manual Energy Selection: 2,5,7,10,20,30,50,70,100,150,200,300,360J
Manual Lockout: Via Pass code
Charge Time: Less than 8 sec.
Analysis Time: 12-16 sec.
Audible Prompts: 13 audible prompts
Visual Prompts: 13 test screen prompts
Controls: Two buttons-On/Off, Discharge 4 software configurable buttons
Output Energy Accuracy: ±10% or 1J (whichever is greater) at 50Ω ±15% or 1J (whichever is greater) at 25-100Ω
Battery
Rechargeable: NiMH 12V, 2.1Ah
Capacity
> 80 discharges at 360 Joules or
> 120 discharges at 200 Joules or
> 150 discharges at 150 Joules or
> 3 hours ECG monitoring
Charge Time
> 1.5 to 2 hours
Non-Rechargeable: Lithium 12V, 5.2 Ah
Capacity
> 200 discharges at 360 Joules or
> 285 discharges at 200 Joules or
> 350 discharges at 150 Joules or
> 6 hours ECG monitoring
> Shelf Life (25°C±15°C)
> 10 years (5 years storage + 5 years standby)
> 5 years standby (after installation)
Capacity may be diminished at operating temperature extremes.