CARDIOVIT AT-101 - High-tech ECG in a compact, budget-priced device.
Performance:
- Intuitive, easy use with direct function keys
- 12-channel Resting ECG
- Graphical LC display
- Storage and serial data transfer for Resting ECG data on PC
- Various formated printouts on internal thermal printer, external DeskJet
or Laser printer
- Memory (optional)
- Measurements
- Interpretation (optional)
- Thrombolysis (optional)
Features:
- Integrated 3-channel monitor
- Alphanumeric keyboard
- Direct keys for the most important functions
- Easy Print: Integrated high-resolution thermal printer, subsequent
printout on thermoreactive paper possible
- Interface for external printer for A4 printouts on regular paper
- Interface for data-transmission (with memory option)
- Capable of data transfer to SCHILLER SEMA Data Management System.
Specifications
System
Dimensions: 290 x 198 x 76 mm, approx. 2.6 kg
Built-in monitor: 76 x 57 mm effective display area, 320 x 240 dots
resolution
On-screen status indicators: Battery status, date, time, power source
Control panel and keyboard: User-friendly, alphanumeric keyboard, LED
indicator and LC display
Power supply requirements: 220 – 240 V (nominal), 50/60 Hz; 110 – 115 V
(nominal), 50/60 Hz; stand-alone operation with built-in rechargeable battery;
LED indicator for mains operation, integrated power supply unit
Battery capacity: 4 hours of normal use
Power consumption: Max. 28 VA
Line frequency filter: Distortion-free suppression of superimposed 50
or 60 Hz sinusoidal interferences by means of adaptive digital filtering
(SCHILLER Powerline Filter SPF)
Frequency range of digital recording system: 0 – 150 Hz (IEC/AHA)
Chart paper: Thermo-reactive, Z-folded, 72 mm wide
Printing process: High-resolution thermal head printer, 8 dots/mm
(amplitude axis), 40 dots/mm (time axis) @ 25 mm/s
Interfaces: RS-232 interface; data transmission to PC (SEMA-200) and
external modem connection; parallel port for external printer
Memory: Possibility to save up to 40 Resting ECG measurements (Option)
Environmental conditions:
> Temperature, operating: 10° to 40° C
> Temperature, storage: -10° to 50° C
> Relative humidity: 25 to 95 % (no condensation)
> Pressure during operation: 700 to 1060 hPa
Safety Standards:
Safety standard: CF according IEC 60601-1 and IEC 60601-2-25
Protection class: I according to IEC 60601-1 (with internal battery)
Classification: IIa according to EEC directive 93/42 (for medical devices)
Conformity: -0123 according to 93/42/EEC; UL 2601-1, C22.2 No 601.1-M90
Technical Data for ECG:
Patient input circuit: Fully floating and isolated, defibrillation protected
(only with original SCHILLER patient cable)
Monitor display:
> 3-channel display of the selected leads
> Heart Frequency, HF
> 25, 50 mm/s
> 5, 10, 20 mm/mV
> Filter status (on/off)
> Insufficient electrode contact
Leads: 12 simultaneous leads: Standard / Cabrera
Chart printout speed: 5/10/25/50 mm/s (manual print)
Sensitivities: 5/10/20 mm/mV, either automatically adjusted or manually
selected
Automatic lead programs:
> 3/6/12-channel presentations of 12 simultaneously recorded leads
> Numerous printout formats can be selected
Data record:
> Patient data (name, age, height, weight, BP), user ID
> Listing of all ECG recording conditions (date, time, filter), (optional)
> Measurement program: ECG measurements results (intervals, amplitudes,
electrical axes), average complexes with measurement reference markings
> Interpretation (C, optional) with adult and pediatric guidance
> Thrombolyse Software (only in combination with interpretive option)
Recording track: 3/6/12-channel presentation, optimal positioning on a
width of 72 mm (internal printer) or A4 (external printer), automatic baseline
adjustment
Filter: Myogram filter (muscle tremor filter): 25 or 35 Hz, SCHILLER
Baseline Stabilizer SBS, SCHILLER Smoothing Filter SSF
ECG amplifier:
> Simultaneous recording of all 9 active electrode signals (= 12 leads)
> Sampling frequency: 1000 Hz
> Pacemaker detection: ≥ ± 2 mV/≥ 0.1 ms
Scope of delivery for CARDIOVIT AT-101 Standard:
CARDIOVIT AT-101 Resting ECG with 12 simultaneous leads, pacemaker detection, automatic ECG measurement
Accessories:
> 1 10-lead patient cable
> 1 set of electrodes or disposable electrodes
> 1 power cable
> 1 pack chart paper
> Operating manual
Software options:
> Thrombolysis Software
> Memory for up to 40 measurements
> Computer-aided ECG interpretation for pediatric and adult ECGs (C)
> SEMA-200 PC software to save, validate and archieve ECG data on PC
Hardware options:
> Equipment trolley
> External printer
*Technical data are subject to change without notice.