The diagnostic value of spirometry as a tool
for the early detection of chronic-obstructive
pulmonary function restriction and other
heart diseases is high. However, with the
constant rise in the number of patients there
is clearly a need to conduct pulmonary
function tests on a routine basis.
SCHILLER combined the two most frequently
used cardiac/pulmonary function tests in
one single unit: the CARDIOVIT AT-2 plus offers ECG and
spirometry all in one product.
This innovative device can be used for quick
and easy initial clarification, for routine
examinations of affected patients, and to
monitor the progress of therapy. This "double
winner" scores high, due to its low purchasing
costs, its economic operation and
the possibility of "double" reimboursement
(for ECG as well as spirometry).
World-renowned ECG technology
- 6-channel print-out in realtime at 5, 25
or 50 mm/s
- Automatic print-out function for 6 and
12 channels on one or two sides of
the paper
- Fast, reliable, computer-aided ECG
interpretation program including complete
measurement for adult and
pediatric ECGs (optional).
Simple to handle – double the benefit!
- Two tests in one – meaning double
reimbursement from health insurance
schemes
- Two different instrument functions –
but only one operating system
- Easy to transport due to low weight
and compact dimensions
- Maximum quality and absolute
reliability
- The built-in printer supplies A4
print-outs in ready-to-file cardex format
- Print-out with diagnosis guidance
and all measurements in less than
20 seconds
- Integrated rechargeable battery
- Expanded print-out possibilities and
unlimited copying function
- Splash-resistant alphanumeric
keyboard with direct function keys
makes it quick and easy to enter
patient data
- Possibility of saving test results for
more than 60 patients (optional)
- Interface enables data transmission
- The curve quality of all 12 leads can
easily be checked on the integrated
3-channel monitor – saving both time
and paper!
- The repetition rate is cut to zero because
the curve quality on the print-out is
optimized using the SSF SCHILLER
Smoothing Filter and the SBS SCHILLER
Baseline Stabilizer.
The very highest standards in spirometry
- The unique SP-250 disposable sensor
minimizes the risk of cross-contamination
- Compact size and low weight
- Calibration is simple and fast
- Inspiratory / expiratory tests: FVC, SVC,
MV, MVV as well as pre and post
medication tests
- Predicted value tables are stored for free
selection.
- A4 print-out within seconds
- Visual patient support due to realtime
curves and measurement information
on the large screen.
Specifications
System
Dimensions: 400 x 330 x 101 mm; approximately 5 kg
Built-in monitor:
- 120 x 90 mm effective display area
- 320 x 240 dots resolution
On-screen status indicators:
- Battery status
- Date, time
- Power source
Control panel and keyboard:
User-friendly, alphanumeric keyboard, LED indicator and
monitor
Power supply requirements: 220 - 240 V (nominal),
50/60 Hz; 110 - 115V (nominal), 50/60 Hz; stand-alone operation
with built-in rechargeable battery. LED indicator for mains
operation, integrated power supply unit.
Battery capacity: Up to 3 hours of normal use
Printer
Chart paper: Thermo-reactive, Z-folded, 210 mm wide (A4),
approx. 60 m, ready to file
Printing process: High-resolution thermal head printer,
8 dots / mm (amplitude axis), 40 dots/mm (time axis)
@ 25 mm/s.
Chart print-out speed: 5 / 10 / 25 / 50 mm/s (manual printout)
Sensitivity: 5 / 10 / 20 mm/m, either automatically adjusted or
manually selected
Recording tracks: 6/12-channel presentation, optimal
positioning on a width of 200 mm, automatic baseline
adjustment
Automatic lead programs:
- 6/12-channel presentations of 12 simultaneously
recorded standard leads on one or more A4 pages
- Numerous print-out formats can be selected
Interface: RS-232 interface to connect spirometry sensor, data
transmission to PC (SEMA-200), and external modem
connection.
Safety standards
Safety standard:
CF according to IEC 60601-1 and IEC 60601-2-25
Protection class: I according to IEC 60601-1 (with internal
battery), Ila according to MDD 93/42 EEC (medical protection
class)
Conformity: according to 93/42 EEC
Environmental conditions:
- Temperature during operation: 10° to 40° C
- Temperature during storage: -10° to 50° C
- Relative humidity: 25 to 95 % (non-condensing)
- Pressure during operation: 700 to 1060 pHa
ECG
Patient input circuit: Fully floating and isolated, defibrillationprotected
(only with original SCHILLER patient cable)
Monitor display:
- 3-channel display of the selected leads
- 25, 50 mm/s
- 5, 10, 20 mm/mV
- Filter status (on/off)
- Insufficient electrode contact
- Heart Frequency, HF
- mm/mv, mm/s
ECG amplifier:
- Simultaneous recording of all 9 active electrode
signals (= 12 leads)
- Sampling frequency: 1000 Hz
- Pacemaker detection: ± 2 mV/0.1 ms
Leads: 12 simultaneous leads: Standard / Cabrera
Data record:
- Patient data (name, age, height, weight, BP), user ID
- Listing of all ECG recording conditions (date, time, filter)
- With optional measurement (M) and interpretation (C)
program: ECG measurement results (intervals, amplitudes,
electrical axes), average complexes with optional
measurement reference markings, guidance on interpreting
adult and pediatric ECGs
Filter
Myogram filter (muscle tremor filter): 25 or 35 Hz, can be
switched on/off
Line frequency filter: Distortion-free suppression of superimposed
50 Hz or 60 Hz sinusoidal interference by means of adaptive
digital filtering.
Frequency range of digital recording system: 0.05 Hz to 150 Hz
(IEC/AHA)
Spirometry (Option)
Measured values:
FVC: FVC, FEV0.5, FEV1.0, FEV3.0, FEV0.5 /FVC, FEV1.0 /FVC,
FEV3.0/FVC, FEF0.2-1.2, FEF25-75%, FEF75-85%, PEF, FEF25%, FEF50%,
FEF75%, FIVC, FIV1.0, FIV1.0/FIVC, FIV1.0/FVC, PIF, FIF50%, FMFT
SVC: SVC, ERV, IRV, TV
MVV: MVV, RR, TV
Presentation possibilities (printout and screen):
- Flow/volume graph
- Volume/time graph
- Table of measured values
- Realtime flow curve
Data record:
- Patient data (name, age, height, weight), user ID
- Registration conditions (date, time, date of last calibration)
- Flow/volume graph and/or volume/time graph
- Table of measured values with PREDICTED/ACTUAL/
DIFFERENTIAL values
- Diagnosis guidance
- Memory for more than 60 ECGs or spirometry records
Prediction equation:
Adults: ECCS, Austria, Crapo, Morris, Knudson, Knudson 76,
Polgar, Berglund, Finland, India, Composite
Children: Quanjer & Tammeling, Austria, India, Knudson,
Knudson 76, Polgar
Extrapolated predicted values
Comparison of PRE/POST medication is possible
Standards Compliance: ATS, OSHA, NIOSH
SPIROVIT SP-250 Pneumotach Flowsensor for pulmonary function
testing with disposable mouthpiece:
Dimensions of SP-250: 118 x 36 x 28 mm; approx. 120 g; 4.6 x
1.4 x 1.1 in, approx. 0.26 Ibs
Measuring method: Pneumotachometer
Measuring accuracy: According ATS < 3 %
Flow impedance: < 0.2 mbar*s/l at 12 l/s
SPIROVIT SP-260 Pneumotach Flowsensor for pulmonary function
testing with reusable mouthpiece:
Dimensions of SP-260: 125 x 36 x 28 mm; approx. 160 g;
4.9 x 1.4 x 1.1 in, approx. 0.34 Ibs
Measuring method, Measuring accuracy, Flow impedance
same as SP-250
Scope of delivery:
- 1 pneumotach sensor; choice of SP-250 disposable
(P/N 2.100022) or SP-260 reusable sensor (P/N 2.100551)
- Accessories: 2 noseclips and either 1 pack of disposable
plastic mouthpieces for SP-250 (P/N 2.100077) or 1 pack of
disposable filters for SP-260 (P/N 2.100123)
- 1 operating manual
Scope of delivery for CARDIOVIT AT-2 plus Standard
Resting ECG with 12 simultaneous leads, pacemaker detection
Accessories:
- 1 10-lead patient cable
- 1 set of electrodes or disposable electrodes
- 1 power cable
- 1 pack chart paper
- 1 operating manual
Software option:
- Automatic ECG measurement and computer-aided
ECG interpretation for pediatric and adult ECGs
- Memory for up to 40 ECGs or 40 spirometry records
- SEMA 200 PC software to save, validate and archive
ECG and spirometry data on a PC