Product Specs
Display
> Size - Dash 3000 – 8.5 in., Dash 4000 – 10.4 in. (diagonal)
> Type - Active-matrix color LCD
> Resolution - 640 by 480 pixels
> Number of traces - 6 (maximum)
> Number of seconds/trace - 4.9 at 25 mm/sec
> Sweep speed - 6.25, 12.5, 25 mm/sec (with erase bar)
Controls
> TrimKnob® control
> Five hard keys - Silence Alarm, Graph Go/Stop, NBP Go/Stop, Zero All and
Power On/Off
> Remote control option - Available
Alarms
> Categories - Patient status and system status
> Priority - 4 levels – Crisis, Warning, Advisory and Message
> Notification - Audible and visual
> Setting - Default and individual
> Silencing - 1 minute, current alarm only
> Pause - 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode and 5
minute, 15 minute, or permanent pause in OR mode
> Volume - Default 70 dB measured at 1 meter
Invasive Blood Pressure
> Number of channels - 1 to 4 (optional); 3rd and 4th invasive lines only
available with Masimo® SpO2
> Transducer sites - Arterial, femoral artery, pulmonary arterial, central
venous, right atrial, left atrial, intracranial and special
> Transducer requirements - Excitation voltage: 5 V dc ± 0.1%
> Transducer output - 5 μV/V/mmHg
Input Specs
> Range - -25 mmHg to 300 mmHg
> Offset - ± 150 mmHg
Output Specs
> Frequency response - dc to 50 Hz (-3 dB)
> Zero balance range - ± 150 mmHg
> Zero balance accuracy - ± 1 mmHg
> Zero balance drift - ± 1 mmHg over 24 hours
> Accuracy - ± 2% or ± 1 mmHg, whichever is greater (exclusive of transducer)
> Alarms - User-selectable upper and lower limits for systolic, diastolic and
mean pressures
ECG
> Standard leads available - I, II, III, V, aVR, aVL and aVF
> 3, 5 and 10 leadwire cable - I, II, III, V, aVR, aVL, aVF, V2, V3, V4, V5
and V6
> Leads analyzed simultaneously - I, II, III and V (multi-lead mode)
> Lead fail - Identifies failed lead
> Alarms - User-selectable upper and lower heart rate limits
Input Specs
> Voltage range - ± 0.5 mV to ± 5 mV
> Signal width - 40 ms to 120 ms (Q to S)
> Heart rate range 3- 0 to 300 bpm
> Input impedance -
Common mode > 10 MΩ at 50/60 Hz
Differential > 2.5 MΩ from dc to 60 Hz
Common mode rejection 90 dB minimum at 50 or 60 Hz
Output Specs
> Impulse response -
For an impulse of 3 mV applied for 100 ms
Displacement following impulse < 0.1 mV
Slope following impulse < 0.3 mV/s
> Frequency response -
Response of non-permanent displays is limited by resolution
to 40 Hz (-3 dB) @25 mm/s. Specified upper frequency
limits may vary by ± 2 Hz.
> Diagnostic mode - 0.67 Hz (+0.4 dB) to 100 Hz (-3 dB)
> For compliance with China National Standard - 1.0 Hz (+0.4dB) to 75 Hz (-3
dB)
> Monitoring mode - 0.67 (+0.4 dB) to 40 Hz (-3 dB)
> Moderate mode - 0.67 (+0.4 dB) to 25 Hz (-3 dB)
> Maximum mode - 5.0 Hz (-0.3 dB) to 25 Hz (-3 dB)
> Noise - < 30 μV (referred to input)
Pacemaker detection/rejection
> Input voltage range - ± 2 mV to ± 700 mV
> Input pulse width - 0.1 ms to 2 ms
> Rise time - 10 μs to 100 μs
> Over/under shoot - 2 mV (max)
> Baseline drift - < 0.5 mV per hour with a ± 700 mV, 2 ms
> Pacemaker pulse - Applied<
Respiration
> Measurement technique - Impedance variation detection
> Range - 0-200 breaths per minute for variations of 1.0 – 10.0 Ω
> Respiration rate - 0-200 breaths per minute
> Base impedance - 100-1000 Ω at 52.6 kHz
> Detection sensitivity - 0.4 to 10 Ω variation
> Waveform display bandwidth - 0.1 to 1.8 Hz (-3 dB)
> Alarms - User-selectable upper and lower respiration rate limits, and
user-selectable apnea limit
Temperature
> Number of channels - 2
Input Specs
> Probe type - YSI Series 400 or 700 (determined by input cable)
> Temperature range - 0°C to 45°C (32°F to 113°F)
> Resolution - ± 0.1°C
Output Specs
> Parameters displayed - T1, T2
> Accuracy (independent of source) - ± 0.1°C for YSI series 400
> Probes - ± 0.3°C for YSI series 700 probes
> Alarms - User-selectable upper and lower limits for T1, T2
Cardiac Output
Input Specs
> Probe type - In-line or bath probe
> Catheter size - 5F, 6F, 7F, 7.5F and 8F
> Injectate volume - 3, 5 or 10 cc
Output Specs
> Parameters displayed - Cardiac output, blood temperature, injectate
temperature and trial number
Range
> Cardiac output - 0.2-15 (liters per minute)
> Blood temperature - 30-42°C
> Injectate temperature - 0-30°C
Accuracy
> Cardiac output - ± 5%
> Blood temperature - ± 0.2°C
> Injectate temperature - ± 0.3°C
> Frequency response - dc to 15 Hz ± 2 Hz
Pulse Oximetry
> Parameters monitored - Arterial oxygen saturation (SpO2) and peripheral
pulse rate (PPR)
> SpO2 range - Nellcor®-compatible 50-100%; Masimo 30-100%
> PPR range - Nellcor-compatible 30-300 BPM; Masimo 25-240 BPM
> Accuracy - Actual accuracy depends on probe. Please reference manufacturer’s
specifications.
> Nellcor compatible SpO2 - ± 2% (70-100% SpO2); ± 3% (50-69% SpO2)
> Masimo SpO2 - ± 2% Adults/Pediatric (70-100% SpO2); SpO2 ± 3% Neonates ≤ 69%
unspecified
> PPR - ± 3 beats per minute
> Alarms - User-selectable upper and lower limits for SpO2 and PPR
DINAMAP® Non-Invasive Blood Pressure
> Technology - DINAMAP classic and SuperSTAT™ (SuperSTAT only available with
Masimo SpO2)
> Measurement technique - Oscillometric
> Displayed parameters - Systolic, diastolic and mean pressures, time of last
measurement
> Measurement modes - Adult ICU and OR modes; manual, auto and stat, neonatal
mode; manual and auto
Systolic
> Adult - 30-285 mmHg
> Pediatric - 30-235 mmHg
> Neonate - 30-140 mmHg
Map
> Adult - 20-260 mmHg
> Pediatric - 20-220 mmHg
> Neonate - 20-125 mmHg
Diastolic
> Adult - 10-220 mmHg
> Pediatric - 10-210 mmHg
> Neonate - 10-110 mmHg
Pulse rate, as displayed in tabular trends
> Adult - 30-200 bpm
> Pediatric - 30-200 bpm
> Neonate - 30-220 bpm
Other Specs
> Overall system accuracy - Meets or exceeds SP 10-1992 AAMI standards
> Automatic cycle times - 0-8 hours
> Tubing length - 12 feet adult, 8 feet neonatal
> Automatic cuff deflation -
Cycle time exceeding 3 minutes (90 seconds neonatal), French mode – Cycle time
exceeding 2 minutes (60 secondsneonatal), power off, or cuff pressure exceeds
294 mmHg (± 6 mmHg) adult, 250 (± 5 mmHg) pediatric, 147 (± 3 mmHg) neonatal
> Cuff sizes - Thigh, large adult, adult, small adult, child, infant and
neonatal, sizes #5 - #1 and assorted long sizes
> Alarms - User-selectable upper and lower limits for systolic, diastolic and
mean pressures
CO2
Technology
> Supports Novametrix CapnoStat (mainstream) and LoFlo (low-flow sidestream)
CO2 technologies
> Principle of operation - Non-dispersive infrared (NDIR) single beam optics,
dual wavelength and no moving parts
> Warm-up time - 2 minutes warm-up time to meet accuracy specifications;
waveform immediate upon power up, calculated end tidal after two breaths
> Cable length (mainstream) - 8 foot (2.4 m)
> Sample line length (low-flow sidestream) - 7 foot (2.1 m)
Information Displayed
> Inspired and expired CO2 concentrations in %, mmHg or kPa; respiratory rate,
continuous CO2 waveform
Measurement range (at 760 mmHg at an ambient temperature of 25°C)
> 0-100 mmHg, 0-13%, 0-12.5 kPa
> PiCO2 /FiCO2 - 0-50 mmHg, 0-6.5%, 0-6.25 kPa
> Respiration rate range - Low-flow SS 0-150 breaths/min
> Mainstream - 0-120 breaths/min
Accuracy (at 760 mmHg at an ambient temperature of 25°C)
> MS -
±2 mmHg or 5%, whichever is greater
SS 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of
reading; all specifications ± 12% of actual from 80-150 BrPM
> Display resolution - 1 mmHg
> Rise time - Less than 200 ms (low-flow sidestream); less than 60 ms
(mainstream adult reusable or SPU); less than 50 ms (mainstream infant
reusable or SPU)
> Respiration rate accuracy - ± 1 breath/min
Compensations
> Automatic barometric pressure - ± 25 mmHg from 530-785 mmHg
> Operator-selectable O2/N2O compensation
Calibration
> Mainstream -
No routine user calibration required
15 second airway adapter zero performed when changing to a different style of
airway adapter
> Low-flow sidestream - No routine user calibration required
Airway adapters and sample lines – mainstream (airway adapters)
> Types - Adult reusable (standard), adult disposable, infant
> Deadspace -
Adult reusable/disposable < 5 cc
Infant disposable < 1 cc
Taper meets ISO 5356-1
Low-flow sidestream airway adapters
> Types - Adult reusable (standard), adult disposable, infant
> Deadspace -
Adult reusable/disposable < 7.3 cc
Infant disposable < 1 cc
Low-flow sidestream sample lines
> Adult, pediatric and infant - Nasal CO2 and nasal CO2/O2
> Adult and pediatric - Nasal/oral CO2 and nasal/oral CO2/O2
Alarms
> CO2 - High inspired CO2; high/low expired CO2
> Respiratory rate Adjustable high and low
Paper Recorder
> Method - Thermal dot array
> Horizontal resolution - 480 dots/in. @25 mm/sec.
> Vertical resolution - 200 dots/in.
> Number of waveform channels - four
> Paper width - 50 mm (1.97 in.)
> Paper length - 30 m (100 ft.)
> Paper speed - 0.1, 0.5, 1, 5, 10, 12.5, 25 and 50 mm/sec. (± 2%)
Analog Output
ECG
> Gain - 1 V/mV ± 10%
> DC offset - ± 100 mV (max)
> Noise - < 5 mV peak to peak 0-300 Hz
> Frequency response - Refer to Frequency Response section under ECG
Blood pressure
> Gain - 10 mV/mmHg ± 2%
> DC offset - ± 20 mV (max)
> Noise - < 5 mV peak to peak 0-300 Hz
> Frequency response - dc to 50 Hz-0/+2 Hz
Battery
> Battery type - Exchangeable Lithium-Ion
> Maximum number of batteries - 2
> Voltage - 11.1 V (nominal)
> Capacity - ≥ 3.45 Ah (varies with manufacturers)
> Charge time - Less than 4 hours each
> Run time - 4 to 5 hours
> Battery life - 500 cycles to 50% capacity
Wireless LAN
> Operating frequency - 2.4 to 2.5 GHz
> Transmit power - 100 mW
> Data rate - 1Mbps and 2Mbps per channel
> Radio technology - Frequency-hopping spread spectrum
> Communication protocol - IEEE 802.11
> UL 1950 Listed (ITE 9B97), CE Mark RF Standard
> US - FCC Part 15 Class B, Europe: ETS 300 328 and ETS 300 826
Environmental Specifications
> Power requirements -
90-132 VAC 50/60 Hz 2.0A
190-264 VAC 50/60 Hz 1.0A
> Power consumption - 75 W (fully loaded)
> Cooling - Convection
> Heat dissipation - 240 Btu/hr. (max)
Operating Conditions
> Ambient temperature - 0-40°C (32-104°F)
> While charging batteries - 0-35°C (32-95°F)
> CO2 sensor - 10-40°C (50-104°F)
> Relative humidity - 5-95% @40°C
> Vibration - MIL-STD 810E, Method 514.4, Category 1
> Altitude - -273 m to 2,943 m (-896 to 9,655 ft.)
Storage Conditions
Do not exceed
> Maximum - 70°C (158°F) at 95% relative humidity
> Minimum - -40°C (-40°F)
> CO2 sensor - -30 to 65°C (-22 to 149°F)
> Batteries - -20 to 60°C (-4 to 140°F)
Physical Specifications
> Height - 26 cm (10.25 in.)
> Depth - 20 cm (8 in.)
> Width - 28 cm (11.0 in.)
> Weight - 5.08 kg (11.2 lbs.)
Warranty
> Standard warranty is one year.