ECG leads
10 lead 12 channel simultaneous 10 second resting ECG
Dimension
296 (w) x 305.5 (h) x 92.5 (d)mm
Approximately 3.5kg
Recording channel
3, 6, 12 channel and 1 channel (60 sec)
Sensitivity
5, 10, 20, auto (I~aVF: 10, V1~V6: 5) mm/mV
Printing speed
12.5, 25, 50 mm/s
Filters
AC (50/60Hz, - 20dB or better)
Muscle (25~35Hz, -3dB or better)
Base line drift (0.1Hz, -3dB or better)
Low pass filter: off, 40Hz, 100Hz, 150 Hz
Display
320 x 240 character LCD display (3 leads trends)
Monitor
Display ID, date, sensitivity, speed, filter, form, rhythm lead
User Interface
Touch screen (alphanumeric and symbol available), keyboard, rotary push-knob (pop-up menu)
Patient data
ID, name, age, sex, height, weight
Basic measurement
Heart rate, PR, QRS, QT/QTc, P-R-T axis
Recorder
Thermal print head, thermal roll paper
Report papers
- Width: A4: 210mm or 8.5"
- Length: A4: 300mm or 11"
Effective writing width: 204mm
Resolution
- Vertical: 8 dot/mm
- Horizontal: 16 dot/mm
Electrical
Internal noise : 20uV (p-p )max
Input circuit : Floating input
Isolated and defibrillation protected
Input impedance : =10MO
Input range : =±5mV
Common Mode Rejection : =+100dB
DC offset voltage : =±300mV
Time constant : 3.2sec
Patient leakage current : <10uA
Frequency response 0.05~150Hz
Signal quality control
Disconnect lead detection
Power
Power supply: AC or built-in battery (option)
100-240 vac, 50/60Hz, 1.0-0.5 A, 60W max
Battery
(Ni-MH) 1 hour of normal use (approx. 100 ECG printouts)
Communication
PC connection with RS-232 interface and LAN
Safety
Class 1, Type BF
Conformity
CE, CSA
Environment
Operating humidity: 30~85%
Operating temperature: 10° ~ 40° C
Atmospheric pressure: 70 ~ 106 KPa
Standard accessory
Power Cord - 1ea.
Patient Cable - 1 ea.
Electrode - 10 ea.
Chart paper - 1 ea.
Operation manual - 1 ea.