Displays, Indicators &
Keys:
SpO2 LED numeric display: 0.43 inch (11 mm) high
Pulse Rate LED numeric display: 0.3 inch (8 mm) high
Pulse Strength: Logarithmically scaled bar graph
Low Battery : LED flashes when approximately 30 minutes of battery
life remains
Alarm Silence: LED flashes
or lights to remind the user that audio alarms are disabled
Probe: LED flashes to alert the operator to check the SpO2 probe placement
Memory
Spot checks and trends can be stored for up to 99 patients and 24
hours of run time.
SpO2:
Range: 0-100%
Accuracy: ±2% at 70-100%; ±3% at 50-69%
Averaging: 4, 8, or 16 pulse beat average
Alarm Range:
> High-50%-100% & OFF (1% steps)
> Low-OFF & 50%-99% (1% steps)
Pulse Rate:
Range: 30-254 bpm
Accuracy: ±2% at 30-254 bpm
Resolution: 1bpm
Alarm Range:
> High - 5-250 & OFF (5 bpm steps)
> Low - OFF & 5-250 (5 bpm steps)
Dimensions:
7.25" H x 3.3" W x 1.85" D
184 mm x 84 mm x 47 mm
Weight:
19 oz. (539 grams)
AC Power Requirements:
105-125V ~ 60Hz
Battery Life:
Standard NiMH (Nickel Metal Hydride) battery:
- Internal rechargeable: not user replaceable
- Fully charges in about 6 hours
- Approximately 24 hours of continuous use
Printer Output:
Data saved every thirty seconds can be printed (trend). Data saved
every four seconds in sleep mode can be output and processed by
commercially available sleep study software.
Environmental Specifications:
Operating Temperature: -4°F to 131°F (-20°C to 55°C)
Storage Temperature: -40°F to 167°F (-40°C to 75°C)
Relative Humidity (non-condensing) : 10-95% (storage), 15-95%
(operating)
Standards:
Certified by TUV to IEC-601-1
CE (EMC), CSA 125, UL544