Displays:
Display: Backlit LCD
Resolution: 240 X 64 pixels
Waveform: Plethysmogram
Graphic Trends: SpO2, Pulse Rate, Pulse Signal Strength
SpO2: SpO2 value, high and low alarm limits
Pulse Rate: Pulse Rate value, high and low alarm limits
Pulse Strength: Logarithmically scaled 8-segment bar graph
Indicators:
Low Battery: Low battery icon when 10 minutes of battery life remain
Audio Status: Yellow LED and displayed symbol indicates alarm audio
paused (2 min.) or off
SpO2 Sensor: Message indicates sensor problem
SpO2 Search: Message indicates monitor is searching for a pulse
Small Pulse: Message indicates oximeter signal strength is low
Charge: Green CHARGE LED indicates external power and battery is charging
Lost Pulse: Message indicates monitor has lost the signal
SpO2:
Range: 0-100%
Accuracy:
> Adult ±2% @70-100% SpO2
> Neonate ±3% @70-100% SpO2
<70% unspecified
Alarm Range:
> High=50-100% & OFF (1% steps)
> Low=50-99% & OFF (1% steps)
Averaging: 4, 8, or 16 second average
Pulse Rate:
Range: 30-254 bpm
Accuracy: ±2 bpm or ±2% (whichever is greater) @ 30-254 bpm
Alarm Range:
> High=30-254 bpm & OFF
> Low=30-254 bpm & OFF
Averaging: 8 or 16 second average
Dimensions:
8"W x 6.5"H x 5"D
(20.3cm x 16.5cm x 12.7cm)
Weight:
4.5lb (2kg)
AC Power Requirements:
Two voltage ranges selected by external switch
90-125VAC 50/60Hz
200-240VAC 50/60Hz
Battery Life:
Standard lead-acid 6 VDC battery:
> Internal rechargeable
> 4 hour continuous use (new)
Printer Output:
Serial infrared link–real time patient data
RS232 trend table output
Environmental Specifications:
Operating Temperature: 32°F to 104°F (0°C to 55°C)
Storage Temperature: -40°F to 167°F (-40° to 75°C)
Relative Humidity (non condensing): 15-95% (operating), 10-95% (storage)
Standards:
EN60601-1 (IEC 601-1)
UL 2601-1 and CAN/CSA 601.1
CAN/CSA NRTL/C